EL3020-URAS26+02SENSOR EL3000 series gas analyzer
EL3020-URAS26+02SENSOR Dynamic Response:
- Warm-Up Time:
- Approx. 30 minutes without thermostat.
- Approx. 2 hours with thermostat.
- 90% Response Time (T90):
- T90 = 2.5 sec under specific conditions:
- Measurement cell length = 175 mm
- Sample gas flow = 60 l/h
- Electronic T90 time (static/dynamic) = 5/0 sec.
- T90 = 2.5 sec under specific conditions:
EL3020-URAS26+02SENSOR Calibration:
- Zero-Point Calibration:
- Using inert gas (e.g., N2) or ambient air free of the sample component.
- End-Point Calibration:
- Utilizing gas-filled calibration cells (optional) or test gas mixtures.
- Recommended to verify calibration cell set values annually.
EL3020-URAS26+02SENSOR Materials in Contact with the Sample Medium:
- Analyzer (Sample Cells):
- Tubing: Aluminum
- Window: CaF2 or BaF2
- Connectors: Stainless steel 1.4305
- Gas Lines and Connectors:
- FPM (Fluorocarbon rubber) hoses, PVDF connectors.
- Option: Stainless steel tubes 1.4571, stainless steel connectors 1.4305
EL3020-URAS26+02SENSOR Gas Inlet Conditions:
- Temperature:
- Sample gas dew point should be at least 5°C below the temperature throughout the sample gas path. Otherwise, a sample gas cooler or condensate trap is necessary.
- Inlet Pressure (pe):
- Range: 2–500 hPa
- Lower pressures require a sample gas pump, while higher pressures require a pressure reducer.
- Outlet Pressure:
- Atmospheric pressure.
- Flow Rate:
- Range: 20–100 l/h.
EL3020-URAS26+02SENSOR Handling Corrosive and Flammable Gases:
- Corrosive Gases:
- Highly corrosive components like chlorine (Cl2) and hydrogen chloride (HCl) require cooling or prior absorption.
- Flammable Gases:
- Version with stainless steel gas lines and connectors suitable for measuring flammable gases in general-purpose environments. Special conditions must be observed (refer to operator’s manual).
For detailed gas connections, please refer to page 16 of the manual.