Description
A6410 Rear Outputs Available Mode Current Outputs 0/4-20 mA output for each channel proportional to main value
For example, RMS or 0 to peak
Open/short circuit proof
Permissible Load <500 Ω
Accuracy ±1% of full scale
Settling Time Configurable, 0 – 10 seconds
A6410 Mode Voltage Outputs 0 – 10 VDC output proportional to main value for each channel
For example, S zero to peak or peak to peak
Open/short circuit proof
Rear Buffered Outputs
Raw buffered output signal, 0 – 12 V peak to peak
Open/short circuit proof
Frequency Range 0.1Hz – 15 kHz (-3 dB)
Permissible Load >10 kΩ
CSI 4500 Interface 4 channel connector designed for CSI 4500 plug & play
The A6410 ModBus and Rack Interface Module is designed for high reliability for the plant’s most critical rotating machinery. It reads parameters from all AMS 6500 modules and outputs these parameters via ModBus TCP/IP and/or ModBus RTU (serial). This 1-slot monitor is used together with other AMS 6500 monitors to build a complete API 670 machinery protection monitor. Applications include steam, gas, compressors and hydro turbo machinery. Operators are provided machinery health parameters in the control environment for seamless integration.
The module powers local graphical displays at the protection rack for machine and instrumentation readout. Either ModBus TCP/IP or ModBus RTU can be configured, or for redundant paths, can be used simultaneously. The AMS 6500 is an integral part of PlantWeb® and AMS software. PlantWeb provides operations integrated machinery health combined with the Ovation® and DeltaV™ process control system. AMS software provides maintenance personnel advanced predictive and performance diagnostic tools to confidently and accurately determine machine malfunctions early.
A6410 Data Output
ModBus RTU (Serial) 19.2 kbaud
Data bits: 8
Parity: Even
Stop bit: 1
Default Address IP address: 192.168.255.1
Subnet mask: 255.255.255.0
ModBus Ethernet RJ 45 socket, 10BaseT
Green LED Power supply OK
Handle Easily remove card and provide plate
for module and sensor identification
Example Data Types Speed, vibration (1/2x, 1x, etc.), gap
value, alarm state, channel OK, phase