EL3040 General Data:

Housing:

  • Version: 19-inch housing (3 height units) for rack-mounting
  • Protection Type: IP20 per EN 60529
  • Materials:
    • Housing: Galvanized sheet steel, outer surfaces varnished
    • Analyzer rear panel: Aluminum, PVC-C
    • Keypad sheet: Polyester
  • Color: Light gray (RAL 7035), basalt gray (RAL 7012)
  • Weight: Approx. 7–15 kg
  • Dimensions: Refer to dimensional drawing on page 13

EL3040 Display and Operation:

  • Display:
    • Backlit graphics display with 240 x 160-pixel resolution
  • Measured Value Display:
    • Numerical value with physical unit, also with bar graph indication
    • Resolution better than 0.2% of the measurement span
    • Simultaneous display of up to 5 measured values
  • Status Display:
    • Symbols in the display; active status messages accessible from measured value display
  • Operation:
    • 5 keys (cursor cross and OK); menu-assisted operation
  • Limit Value Monitoring:
    • Configurable limit values; alarms output via digital ports

Pressure Sensor:

  • Use:
    • Standard for Uras26 and Caldos27, option for Magnos206
  • Materials in Contact with Sample Medium:
    • Silicon gel, plastics, FPM (Fluorocarbon rubber)

EL3040 Electromagnetic Compatibility:

  • Noise Immunity:
    • Tested to EN 61326:1997 + A1:1998 + A2:2001 + A3:2003
  • Emitted Interference:
    • Tested to EN 61326:1997 + A1:1998 + A2:2001 + A3:2003, EN 61000-3-2:2000, and EN 61000-3-3:1995 + A1:2001

EL3040 Electrical Safety:

  • Tested per EN 61010-1:2001
  • Protection Class: I
  • Overload Category/Pollution Level:
    • Power supply: III/2
    • Signal inputs and outputs: III/2
  • Safe Isolation:
    • Galvanically isolated power supply
  • Mechanical Stress:
    • Operation: Vibration and shock tests compliant with standards
    • Transport: Vibration and shock tests compliant with standards

Ambient Conditions:

  • Ambient Temperature:
    • Operation: +5 to +45 °C
    • Uras26 in combination with another analyzer: +5 to +40 °C
    • Storage and transport: –25 to +65 °C
  • Relative Humidity: < 75%, slight condensation allowed
  • Air Circulation: Sufficient air circulation required for multiple housings in a 19-inch rack

EL3040 Power Supply:

  • Input Voltage: 100–240 V AC (– 15 %, + 10 %) 50–60 Hz (± 3 Hz)
  • Power Consumption: Max. 187 W
  • Connection: 3-pin plug per EN 60320-1/C14; connection cable supplied, length 2 .

 

Application scope:

 

Company environment: