Description
EL3040 series gas analyzer
EL3040 General Data:
Housing:
- Version: 19-inch housing (3 height units) for rack-mounting
- Protection Type: IP20 per EN 60529
- Materials:
- Housing: Galvanized sheet steel, outer surfaces varnished
- Analyzer rear panel: Aluminum, PVC-C
- Keypad sheet: Polyester
- Color: Light gray (RAL 7035), basalt gray (RAL 7012)
- Weight: Approx. 7–15 kg
- Dimensions: Refer to dimensional drawing on page 13
EL3040 Display and Operation:
- Display:
- Backlit graphics display with 240 x 160-pixel resolution
- Measured Value Display:
- Numerical value with physical unit, also with bar graph indication
- Resolution better than 0.2% of the measurement span
- Simultaneous display of up to 5 measured values
- Status Display:
- Symbols in the display; active status messages accessible from measured value display
- Operation:
- 5 keys (cursor cross and OK); menu-assisted operation
- Limit Value Monitoring:
- Configurable limit values; alarms output via digital ports
Pressure Sensor:
- Use:
- Standard for Uras26 and Caldos27, option for Magnos206
- Materials in Contact with Sample Medium:
- Silicon gel, plastics, FPM (Fluorocarbon rubber)
EL3040 Electromagnetic Compatibility:
- Noise Immunity:
- Tested to EN 61326:1997 + A1:1998 + A2:2001 + A3:2003
- Emitted Interference:
- Tested to EN 61326:1997 + A1:1998 + A2:2001 + A3:2003, EN 61000-3-2:2000, and EN 61000-3-3:1995 + A1:2001
EL3040 Electrical Safety:
- Tested per EN 61010-1:2001
- Protection Class: I
- Overload Category/Pollution Level:
- Power supply: III/2
- Signal inputs and outputs: III/2
- Safe Isolation:
- Galvanically isolated power supply
- Mechanical Stress:
- Operation: Vibration and shock tests compliant with standards
- Transport: Vibration and shock tests compliant with standards
Ambient Conditions:
- Ambient Temperature:
- Operation: +5 to +45 °C
- Uras26 in combination with another analyzer: +5 to +40 °C
- Storage and transport: –25 to +65 °C
- Relative Humidity: < 75%, slight condensation allowed
- Air Circulation: Sufficient air circulation required for multiple housings in a 19-inch rack
EL3040 Power Supply:
- Input Voltage: 100–240 V AC (– 15 %, + 10 %) 50–60 Hz (± 3 Hz)
- Power Consumption: Max. 187 W
- Connection: 3-pin plug per EN 60320-1/C14; connection cable supplied, length 2 .
Application scope:
Company environment: