PCI-6133 DAQ I/O terminal

¥4,230.00

Input current during
overvoltage conditions: ±20 mA max
Input FIFO size
NI 6132: 16 MS
NI 6133: 16 or 32 MS

Category: SKU: PCI-6133 Tag:
Whatsapp:+86 15359293870
WeChat:+86 18106937731
                E-mail:geabbdcs@gmail.com
Contacts:kelly CHEN

Description

Note:With NI-DAQmx, National Instruments has revised its terminal names so they are easier to understand and more consistent among NI hardware and software products. The revised terminal names used in this document are usually similar to the names they replace. For a complete list of Traditional NI-DAQ terminal names and their NI-DAQmx equivalents, refer to the Terminal Name Equivalents table in the S Series Help.

 

Analog Input
Input Characteristics
Number of channels
NI 6132 …………………………………4
NI 6133 …………………………………8
Type of ADC
Resolution ……………………………..14 bits, 1 in 16,384
Pipeline …………………………………0
Sampling rate
Maximum………………………………2.5 MS/s per channel
Minimum ………………………………No minimum
Input impedance
AI – to AI GND ……………………..100 MΩ in parallel with
10 pF
AI + to AI GND ……………………..100 MΩ in parallel with
10 pF
Input bias current………………………..±2 pA typ, ±25 pA max
Input offset current ……………………..±1 pA typ, ±10 pA max
Input coupling…………………………….DC
Max working voltage for all analog input channels
Positive input (AI +)………………. ±11 V for all ranges,
Measurement Category I
Negative input (AI –) …………….. ±11 V for all ranges,
Measurement Category I

Physical
Dimensions (not including connectors)
NI PCI-6132/6133 …………………. 31.2 cm × 10.6 cm
(12.3 in. × 4.2 in.)
NI PXI-6132/6133…………………. 16.0 cm × 10.0 cm
(6.3 in. × 3.9 in.)
I/O connector ……………………………. 68-pin male SCSI-II type

Environmental
Operating temperature ……………….. 0 to 50 °C
Storage temperature …………………… –20 to 70 °C
Humidity ………………………………….. 10 to 90% RH,
noncondensing
Maximum altitude……………………… 2,000 m
Pollution Degree ……………………….. 2
Indoor use only

 

Electromagnetic Compatibility
The NI 6132/6133 devices are designed to meet the
requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A

 

Note:Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.